Q: Can this kit work on Dynex 2 platform?
A: Generally, as long as the Dynex DS2 can take readings at 450nm and the kit fits in the plate reader, then it should work.
Q: Do you have the template as to how controls, samples need to be set up in the plate?
A: 1) First, bring all reagents and samples to room temperature (18°C-25°C) naturally for 30min before starting assay procedures. DO NOT use hot water baths to thaw samples or reagents. If necessary, doing a low - speed centrifugation for one or two seconds to concentrate the Positive/Negative Controls to the bottom of the vials. The Microelisa Stripplate is detachable, detach unused part strips from the plate frame, return them to the foil pouch with the desiccant pack, and reseal for preventing damps.
2) Set Positive Control wells, Negative Control wells and Sample wells. Add Positive Control 50µl to each Positive Control well, add Negative Control 50µl to each Negative Control well. Add sample 10µl and Sample Diluent 40µl to each Sample wells.
PC: Positive Control; NC: Negative Control; S01-S92: Sample
3) Add 100µl of HRP-conjugate reagent to Positive Control wells, Negative Control wells and Sample wells, cover with an adhesive strip and incubate for 60 minutes at 37°C.
4) Wash the Microtiter Plate 4 times.
5) Add Chromogen Solution A 50µl and Chromogen Solution B 50µl to each well successively. Gently mix and then protect from light to incubate for 15 minutes at 37°C.
6) Add 50µl Stop Solution to each well. The color in the wells should change from blue to yellow. If the color in the wells is green or the color change does not appear uniform, gently tap the plate to ensure thorough mixing.
7) Read the Optical Density (O.D.) at 450 nm using a microtiter plate reader within 15 minutes.